Mdr 745 Specialist

EU MDR 2017/745 compliance specialist for medical device classification, technical documentation, clinical evidence, and post-market surveillance. Covers Ann...

MIT-0 · Free to use, modify, and redistribute. No attribution required.
1 · 1.4k · 5 current installs · 5 all-time installs
byAlireza Rezvani@alirezarezvani
MIT-0
Security Scan
VirusTotalVirusTotal
Benign
View report →
OpenClawOpenClaw
Benign
high confidence
Purpose & Capability
Name and description (EU MDR specialist) match the manifest: SKILL.md contains classification, technical documentation, clinical evidence and PMS guidance, and the included scripts/references implement a gap-analysis and templates appropriate to that purpose. The presence of a local Python 'mdr_gap_analyzer.py' is coherent with a compliance-assessment tool. The only minor provenance issue is that the skill's source/homepage is unknown (no publisher link), but that does not make the requested capabilities disproportionate.
Instruction Scope
SKILL.md instructions are scoped to MDR classification, documentation and evidence workflows. The instructions do not direct the agent to read unrelated system files, request environment variables, or post data to external endpoints. They don't give open-ended permissions ('gather whatever context you need').
Install Mechanism
No install spec (instruction-only) — lowest install risk. A single Python script is included but there is no download-from-URL, package installation, or extract step. Running the script locally would require Python, which is typical and reasonable for this tool.
Credentials
The skill declares no required environment variables, credentials, or config paths. The included code and docs do not reference secrets or external service keys. Requested access is proportional to the stated functionality.
Persistence & Privilege
Flags show default privileges (always: false, agent invocation enabled). The skill does not request permanent/force-included presence and does not appear to modify other skills or system-wide agent settings.
Assessment
This skill appears internally consistent and implements what it claims: MDR guidance, templates, and a local gap-analysis script. Before installing or running it: (1) verify the author's identity/source if you require provenance (there's no homepage listed), (2) review the Python script yourself or run it in a sandboxed environment if you have concerns, (3) do not supply unrelated secrets or system credentials to the skill, and (4) if you will rely on the output for regulatory submissions, have a qualified clinical/regulatory evaluator review the results for completeness and correctness.

Like a lobster shell, security has layers — review code before you run it.

Current versionv2.1.1
Download zip
latestvk9785aqfp10sjxdnchjgw1tbjd82kv5g

License

MIT-0
Free to use, modify, and redistribute. No attribution required.

SKILL.md

MDR 2017/745 Specialist

EU MDR compliance patterns for medical device classification, technical documentation, and clinical evidence.


Table of Contents


Device Classification Workflow

Classify device under MDR Annex VIII:

  1. Identify device duration (transient, short-term, long-term)
  2. Determine invasiveness level (non-invasive, body orifice, surgical)
  3. Assess body system contact (CNS, cardiac, other)
  4. Check if active device (energy dependent)
  5. Apply classification rules 1-22
  6. For software, apply MDCG 2019-11 algorithm
  7. Document classification rationale
  8. Validation: Classification confirmed with Notified Body

Classification Matrix

FactorClass IClass IIaClass IIbClass III
DurationAnyShort-termLong-termLong-term
InvasivenessNon-invasiveBody orificeSurgicalImplantable
SystemAnyNon-criticalCritical organsCNS/cardiac
RiskLowestLow-mediumMedium-highHighest

Software Classification (MDCG 2019-11)

Information UseCondition SeverityClass
Informs decisionNon-seriousIIa
Informs decisionSeriousIIb
Drives/treatsCriticalIII

Classification Examples

Example 1: Absorbable Surgical Suture

  • Rule 8 (implantable, long-term)
  • Duration: > 30 days (absorbed)
  • Contact: General tissue
  • Classification: Class IIb

Example 2: AI Diagnostic Software

  • Rule 11 + MDCG 2019-11
  • Function: Diagnoses serious condition
  • Classification: Class IIb

Example 3: Cardiac Pacemaker

  • Rule 8 (implantable)
  • Contact: Central circulatory system
  • Classification: Class III

Technical Documentation

Prepare technical file per Annex II and III:

  1. Create device description (variants, accessories, intended purpose)
  2. Develop labeling (Article 13 requirements, IFU)
  3. Document design and manufacturing process
  4. Complete GSPR compliance matrix
  5. Prepare benefit-risk analysis
  6. Compile verification and validation evidence
  7. Integrate risk management file (ISO 14971)
  8. Validation: Technical file reviewed for completeness

Technical File Structure

ANNEX II TECHNICAL DOCUMENTATION
├── Device description and UDI-DI
├── Label and instructions for use
├── Design and manufacturing info
├── GSPR compliance matrix
├── Benefit-risk analysis
├── Verification and validation
└── Clinical evaluation report

GSPR Compliance Checklist

RequirementEvidenceStatus
Safe design (GSPR 1-3)Risk management file
Chemical properties (GSPR 10.1)Biocompatibility report
Infection risk (GSPR 10.2)Sterilization validation
Software requirements (GSPR 17)IEC 62304 documentation
Labeling (GSPR 23)Label artwork, IFU

Conformity Assessment Routes

ClassRouteNB Involvement
IAnnex II self-declarationNone
Is/ImAnnex II + IX/XISterile/measuring aspects
IIaAnnex II + IX or XIProduct or QMS
IIbAnnex IX + X or X + XIType exam + production
IIIAnnex IX + XFull QMS + type exam

Clinical Evidence

Develop clinical evidence strategy per Annex XIV:

  1. Define clinical claims and endpoints
  2. Conduct systematic literature search
  3. Appraise clinical data quality
  4. Assess equivalence (technical, biological, clinical)
  5. Identify evidence gaps
  6. Determine if clinical investigation required
  7. Prepare Clinical Evaluation Report (CER)
  8. Validation: CER reviewed by qualified evaluator

Evidence Requirements by Class

ClassMinimum EvidenceInvestigation
IRisk-benefit analysisNot typically required
IIaLiterature + post-marketMay be required
IIbSystematic literature reviewOften required
IIIComprehensive clinical dataRequired (Article 61)

Clinical Evaluation Report Structure

CER CONTENTS
├── Executive summary
├── Device scope and intended purpose
├── Clinical background (state of the art)
├── Literature search methodology
├── Data appraisal and analysis
├── Safety and performance conclusions
├── Benefit-risk determination
└── PMCF plan summary

Qualified Evaluator Requirements

  • Medical degree or equivalent healthcare qualification
  • 4+ years clinical experience in relevant field
  • Training in clinical evaluation methodology
  • Understanding of MDR requirements

Post-Market Surveillance

Establish PMS system per Chapter VII:

  1. Develop PMS plan (Article 84)
  2. Define data collection methods
  3. Establish complaint handling procedures
  4. Create vigilance reporting process
  5. Plan Periodic Safety Update Reports (PSUR)
  6. Integrate with PMCF activities
  7. Define trend analysis and signal detection
  8. Validation: PMS system audited annually

PMS System Components

ComponentRequirementFrequency
PMS PlanArticle 84Maintain current
PSURClass IIa and higherPer class schedule
PMCF PlanAnnex XIV Part BUpdate with CER
PMCF ReportAnnex XIV Part BAnnual (Class III)
VigilanceArticles 87-92As events occur

PSUR Schedule

ClassFrequency
Class IIIAnnual
Class IIb implantableAnnual
Class IIbEvery 2 years
Class IIaWhen necessary

Serious Incident Reporting

TimelineRequirement
2 daysSerious public health threat
10 daysDeath or serious deterioration
15 daysOther serious incidents

EUDAMED and UDI

Implement UDI system per Article 27:

  1. Obtain issuing entity code (GS1, HIBCC, ICCBBA)
  2. Assign UDI-DI to each device variant
  3. Assign UDI-PI (production identifier)
  4. Apply UDI carrier to labels (AIDC + HRI)
  5. Register actor in EUDAMED
  6. Register devices in EUDAMED
  7. Upload certificates when available
  8. Validation: UDI verified on sample labels

EUDAMED Modules

ModuleContentActor
ActorCompany registrationManufacturer, AR
UDI/DeviceDevice and variant dataManufacturer
CertificatesNB certificatesNotified Body
Clinical InvestigationStudy registrationSponsor
VigilanceIncident reportsManufacturer
Market SurveillanceAuthority actionsCompetent Authority

UDI Label Requirements

Required elements per Article 13:

  • UDI-DI (device identifier)
  • UDI-PI (production identifier) for Class II+
  • AIDC format (barcode/RFID)
  • HRI format (human-readable)
  • Manufacturer name and address
  • Lot/serial number
  • Expiration date (if applicable)

Reference Documentation

MDR Classification Guide

references/mdr-classification-guide.md contains:

  • Complete Annex VIII classification rules (Rules 1-22)
  • Software classification per MDCG 2019-11
  • Worked classification examples
  • Conformity assessment route selection

Clinical Evidence Requirements

references/clinical-evidence-requirements.md contains:

  • Clinical evidence framework and hierarchy
  • Literature search methodology
  • Clinical Evaluation Report structure
  • PMCF plan and evaluation report guidance

Technical Documentation Templates

references/technical-documentation-templates.md contains:

  • Annex II and III content requirements
  • Design History File structure
  • GSPR compliance matrix template
  • Declaration of Conformity template
  • Notified Body submission checklist

Tools

MDR Gap Analyzer

# Quick gap analysis
python scripts/mdr_gap_analyzer.py --device "Device Name" --class IIa

# JSON output for integration
python scripts/mdr_gap_analyzer.py --device "Device Name" --class III --output json

# Interactive assessment
python scripts/mdr_gap_analyzer.py --interactive

Analyzes device against MDR requirements, identifies compliance gaps, generates prioritized recommendations.

Output includes:

  • Requirements checklist by category
  • Gap identification with priorities
  • Critical gap highlighting
  • Compliance roadmap recommendations

Notified Body Interface

Selection Criteria

FactorConsiderations
Designation scopeCovers your device type
CapacityTimeline for initial audit
Geographic reachMarkets you need to access
Technical expertiseExperience with your technology
Fee structureTransparency, predictability

Pre-Submission Checklist

  • Technical documentation complete
  • GSPR matrix fully addressed
  • Risk management file current
  • Clinical evaluation report complete
  • QMS (ISO 13485) certified
  • Labeling and IFU finalized
  • Validation: Internal gap assessment complete